Contd.
ACOUSTIC PD INSTRUMENTATION
Acoustic
partial discharge detection set up is very simple, consisting of a sensor,
filter, preamplifier, some type of data acquisition system (e.g., storage
oscilloscope) and some output device.
The detected AE signals from the sensors are preamplified and filtered for
spurious signals. These filtered signals are further amplified and are
passed to a digital storage oscilloscope where the transients are captured
and stored digitally. The stored transient can be transferred to an X-Y
recorder via a digital to analog converter, or passed in the digital form to
a spectrum analyzer.

Alternatively,
the amplified analog signal can bypass the digital storage oscilloscope and
be fed to a data acquisition system and a PC, were the signals can be
analyzed.
The
system frequency response (time constant) determines most of the system
detection characteristics. The sensitivity and signal-to-noise ratio are
determined primarily by the amplitude and frequency characteristics of the
signal that arrives at the sensor and the ambient mechanical back-ground
noise.
The
coupling from the apparatus under test to the sensor should be considered as
an integral part of the system, as it strongly influences system
characteristics.
LOCATION
OF DISCHARGES
In
order to determine the severity of a PD problem, it is necessary to locate
its geometric location as well as its magnitude. The possibility of PD
location is one of the major features of acoustic discharge determination.
Location can be based on either measurement of time of signal arrival at a
sensor or of signal level.
An
arrangement for on-line measurement instrumentation for PD location in case
of transformer is shown in fig. 4. Two AE detection channels are used in
this set up. From the analysis of the difference in signal magnitude and the
time delay of the two signals the location of the PD source is determined.
EXPERIMENTAL
SET UP
A test chamber (transformer tank
model) as shown in fig. 5 was designed and fabricated for the experiment
purpose. Various types of partial discharges can be simulated inside the
chamber by using the electrode system and the PD measurements by both the
conventional electrical methods and acoustic emission methods can be
performed at laboratory. Then the correlation between the test results of
the two methods shall be performed.
CONCLUSION
This paper describes the latest
techniques being developed for the identification of partial discharges
inside the equipment and to analyze the nature of fault, their origin and
location so that appropriate remedial measures can be adopted.